Model 12962A,
12963A, 12964A
Room
Temperature Sample Holders
|
The 12962A sample holder is designed to allow
accurate impedance tests of solid materials to be performed at room
temperature. The sample holder consists of two parallel
electrodes, one of which is fixed in position and the other
which can be moved into contact with the sample by adjustment
of a micrometer. The sample holder makes use of guard
ring and reference techniques in order to improve accuracy by reducing “fringing”
effects at the edge of the sample. The standard electrode
provided is 20mm diameter.
Model 12963A electrode kit
This auxiliary electrode kit for the 12962A sample holder
allows measurements of different sizes of materials. 10, 30
and 40 mm electrodes are supplied in this kit.
Model 12964A liquid sample holder
The liquid sample holder also makes use of guard ring
techniques and is shaped as a container in order to be
able to measure small samples of liquid or powders. The sample holder is
easily dissembled for cleaning.
|
 |
Measurement Techniques
Guard ring
The 12962A, 12963A and 12964A sample holders make use of a guard
ring on the fixed electrode in order to
reduce the effect of stray field lines at the edge of the sample which
would otherwise lead to measurement errors.
The guard ring ensures that the electric field lines are parallel
throughout the part of the sample which contributes
to the impedance measurement.
Normalized Measurements
In order to obtain best accuracy from any sample holder, it is
advisable to make use of normalization measurement techniques. The
sample holder allows normalized measurements to be performed with
ease. Typically measurements are taken with the sample material in
the cell and then repeated without the sample (using an airgap). The
air-gap can easily be set up using the micrometer. The empty cell
measurements can then be used to normalize the sample results to
further reduce errors due to fringing, connections and
instrumentation.
This sample holder is designed for room temperature measurements only,
other sample holders and systems are available for high and low
temperature measurements in cryostats and furnaces.
Specifications
|
Electrode diameter |
10, 20, 30 and 40 mm |
|
Connections |
2-terminal |
|
Sample types |
solid, liquid and powder |
|
Sample thickness |
0.2 mm to 25.4 mm |
|
Sample width and length |
≥ 10mm |
|
Impedance range |
1 ohm to 100 Tohms |
|
|
|
|

Electrode arrangement of he 12962A sample holder |