| 129610A
Cryostat Specification |
|
Cryostat (Special
Model) |
Janis STVP-200
special model with static gas sample chamber |
|
Cryostat construction |
Tubular cryostat with
vacuum shroud and radiation shield |
|
Sample space |
Static gas (e.g.
helium) |
|
|
Inner reservoir - LHe
or LN2 |
|
|
Sample space - Static
gas (He) |
|
Temperature range |
5 K to
600 K - LHe |
|
|
78 K to
600 K - LN2 |
|
Temperature stability |
±0.05 K |
|
Temperature probes |
Type E thermocouple
- Exchange gas heater |
|
Cryogen transfer line |
LHe / LN2
transfer line (included) |
|
Dimensions |
31" (786 mm)
height, 5.75" (146 mm) diameter |
|
Approximate Weight |
25 lb (11.3 kg) |
|
Optional Dewars |
-
A dewar (LN2 or LHe) is needed to operate this
system.
- Compatibility with LN2 dewars that are not
supplied as part of the system cannot be guaranteed. |
|
Option A1 |
LN2 dewar
(10 L) LN-10 |
|
Dimensions |
23"
(584 mm) height, 11.25" (286 mm) diameter |
| Empty
Weight |
12.5 lb
(5.7 kg) |
|
Option A2 |
LN2 dewar
(20 L) LN-20 |
|
Dimensions |
24.6"
(625 mm) height, 15" (381 mm) diameter |
| Empty
Weight |
23 lb
(10.4 kg) |
|
Option A3 |
LN2 dewar
(50 L) LN-10 |
|
Dimensions |
30.5"
(775 mm) height, 18.5" (470 mm) diameter |
| Empty
Weight |
38 lb
(17.2 kg) |
|
|
LN2 dewar
(50 L) LN-10 |
|
Dimensions |
30.5"
(775 mm) height, 18.5" (470 mm) diameter |
| Empty
Weight |
38 lb
(17.2 kg) |
|
Option B1
(on application) |
-
LHe dewars are usually provided by cryogen supplier with
standard 0.5" transfer line fitting. |
|
Option C |
Turbo pump (42 L/s)
TP-75
- or equivalent is needed for system
operation. |
| Temperature
Controller (included) |
|
Controller |
Lake Shore 332S-T2
Dual Loop (sample and heater) |
|
Dimensions |
8.5"
(216 mm) W × 3.5" (89 mm) H ×
14.5" (368 mm) D |
|
Weight |
10.5 lb
(4.8 kg) |
|
Power V AC |
100, 120, 220, 240
(+6%, -10%), 50 or 60 Hz, 150 VA |
|
Interface |
IEEE-488 (GPIB) (PC
requires NI IEEE-488 interface - not included in system
price) |
| Solid Sample
Holder (included) |
|
Electrodes |
Gold-plated brass
electrodes (2x) |
|
Maximum sample size |
0.314"
(8 mm) height, 1.18" (30 mm) diameter |
|
Temperature range |
5 K to
600 K |
|
Lower electrode |
Sliding stage with
screw adjustment. Spring-loaded to allow sample expansion /
contraction |
|
Connections |
4 BNC terminals for 2
or 4 terminal measurements |
|
Internal connections |
High temperature
coaxial |
|
Sample measurement |
Uses guard ring
techniques to reduce "fringing" |
|
Sample heater |
Located close to
sample with isothermal shield |
|
Temperature sensor |
Type E thermocouple |
| Liquid Sample
Holder (included) |
|
Position |
Locates between the
electrodes in the solid sample holder |
|
Holder materials |
Gold-plated brass
electrodes (2x), PEEK body |
|
Temperature range |
5 K to
500 K |
|
Liquid volume |
0.65 mL |
|
Dimensions |
0.315"
(8 mm) high, 1.18" (30 mm) diameter |
| System
Compatibility |
|
The model
129610A Cryostat is fully compatible with all Solartron
Analytical materials test systems |